Evaluation of Medical Equipment/Implant Materials

Related menu

Bendability Test of Stentgrafts

In-page menu

Evaluation of bendability necessary to prevent kinks, etc.

Outline of Measurement Technology

Stentgrafts are medical devices used to implant stents in artificial blood vessels. Since it is important to understand their bendability, bendability properties such as maximum bending load, load per unit angle, etc. are evaluated from the bending angle-bending load curve when one end of a stent graft is held and the other end is rotated along an arc. In case of small diameter stents, a 3-point bending test or cantilever bending test is performed.

The test jigs and test conditions conform to "JIS T 0401 Mechanical testing methods for stentgrafts."

Outline of Test

Test specimen Stentgraft, stent *
Test standard JIS T 0401 "Mechanical testing methods for stentgrafts"
Test environment In atmosphere
Test method Static test
Max. load 250N
Max. bending angle ± 50° *

* Please consult with JFE-TEC.

Schematic Diagram of Stentgraft Bendability Test

  • Bending Test

  • Bending Tests for Small Diameter
    (≤ φ 20) Specimens

Related Links

Inquiry about Products and Services

  • MAIL
  • Flow
  • FAQ